7.2
Processing limitation
Frequent processing can damage this product. Observe the safe-
ty instructions for checking the product. With visible damage or
functional impairment, the product may no longer be used.
7.3
Dismantling
The product must be cleaned and disinfected while it is disman-
tled and the jaws are open.
1. Disconnect the connection cable from the handle (no pic-
ture).
2. Press both release buttons (1) at the same time and pull the
shaft tube out of the handle.
3. Unscrew the insert on the joint sleeve (2) out of the shaft
tube.
4. Pull the insert out of the shaft tube.