6.7.3. "Program" for Continuous Rotary File Systems
For convenience, the device is delivered with 8 programs with default values of torque and
speed (see chapter 13. Program – Individual Continuous Rotary Program).
Press the FILE key to select the next program number.
Press the FILE ⊳ key to select the previous program number.
• To individually change these settings simply "overwrite" them as described below. This
allows you to compile your own instrument sequence independently from file manufacturer
or recommended sequences.
• To return to default settings see chapter 6.8. Factory Default Parameters.
• For recording your individual settings, see table in chapter 13. Program – Individual
Continuous Rotary Program.
6.7.4. Changing Speed and Torque
NOTES
• Speed and torque cannot be changed for reciprocating systems.
• While the motor handpiece is in motion, speed and torque can be changed but they cannot
be saved.
When the desired continuous rotary file is selected, press the + or – SPEED keys to select the
desired speed setting.
When the speed value is changed from the default setting, SPEED is displayed between
brackets.
If the MEMO key is not pressed to save the setting, the setting will be lost once a different file
setting is selected.
The speed setting can be adjusted from 250 to 1000 rpm in increments of 50 rpm and from
1000 to 1200 rpm in increments of 100 rpm.
Press the + or – TORQUE keys to select the desired torque setting.
When the torque value is changed from the default setting, TORQUE is displayed between
brackets.
If the MEMO key is not pressed to save the setting, the setting will be lost once a different file
setting is selected.