9 RECOMMENDED TEST PRESSURES & ACCEPTABLE RANGES (continued)
Megapascals (MPa)
Gauge
Elcometer AVU
Pressure
Acceptable Range
5.0
4.25 - 5.75
10.0
8.50 - 11.50
15.0
12.75 - 17.25
Megapascals (MPa)
Elcometer AVU
Gauge
Pressure
Acceptable Range
0.5
2.74 - 3.71
1.0
5.48 - 7.42
1.5
8.23 - 11.13
2.0
10.97 - 14.84
Elcometer AVU reading is multiplied by the conversion factor of 6.4516 to give actual measured result.
9
ELCOMETER 106 SCALE 4 (Accuracy: ±15% of the reading)
Gauge
Elcometer AVU
Pressure
Acceptable Range
20.0
17.00 - 23.00
22.0
18.70 - 25.30
ELCOMETER 106 SCALE 6 (Accuracy: ±15% of the reading)
Elcometer AVU
Gauge
Pressure
Acceptable Range
2.5
13.71 - 18.55
3.0
16.45 - 22.26
3.5
19.19 - 25.97
Pounds Per Square Inch (psi)
Gauge
Elcometer AVU
Pressure
Acceptable Range
500
425 - 575
1000
850 - 1150
1500
1275 - 1725
2000
1700 - 2300
Pounds Per Square Inch (psi)
Elcometer AVU
Gauge
Pressure
Acceptable Range
50
274 - 371
100
548 - 741
150
822 - 1112
200
1096 - 1483
250
1371 - 1854
Gauge
Elcometer AVU
Pressure
Acceptable Range
2500
2125 - 2875
3000
2550 - 3450
3200
2720 - 3680
Elcometer AVU
Gauge
Pressure
Acceptable Range
300
1645 - 2225
350
1919 - 2596
400
2193 - 2967
450
2467 - 3338
500
2742 - 3710