MIL'S HOSPIVAC 3 E25 Instrucciones Generales De Puesta En Funcionamiento Y Mantenimiento página 18

Unidad de vacío cabine con procom 2 o cyclic 2020
Tabla de contenido

Publicidad

INSTRUCCIONES GENERALES DE PUESTA EN FUNCIONAMIENTO
HOSPIVAC CABINE con PROCOM 2 o CYCLIC 2020
Guidance and manufacturer's declaration – electromagnetic immunity - for ME equipment and ME systems that are not
The vacuum units with PROCOM2 or CYCLIC 2020 automate is intended for use in the electromagnetic environment specified
below. The customer or the user of the vacuum units with PROCOM2 or CYCLIC 2020 automate should assure that it is used
in such an electromagnetic environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
(1) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the vacuum units with PROCOM2 or CYCLIC 2020 automate is used exceeds
the applicable RF compliance level above, the vacuum units with PROCOM2 or CYCLIC 2020 automate should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the vacuum units with PROCOM2 or CYCLIC 2020 automate.
(2) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Pàgina 17/18
06/2015
life-supporting
IEC 60601 test level
Compliance level
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
Electromagnetic environment – guidance
Portable
equipment should be used no closer to any part
of the vacuum units with PROCOM2 or CYCLIC
2020 automate, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1.16 √P
3 Vrms
d = 1.16 √P de 80 MHz à 800 MHz
3 V/m
d = 2.33 √P de 800 MHz à 2,5 GHz
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
(1) should be less than the compliance level in
each frequency range. (2)
Interference may occur in the vicinity of
equipment marked with the following symbol:
Manual n° 517391 versión 5
and
mobile
RF
communications
manufacturer
and
is
d
the

Publicidad

Tabla de contenido
loading