Trunnion Ring - Philips SRO 2550 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

If a not compatible collimator is used
The alignment ring is mounted when the tube is delivered.
Remove the alignment ring and dispose it.
It is not recyclable.
Proceed the installation in the common way.
See
1. Coupling flange
figure 4
2. Alignment ring
5.1.2

Trunnion ring

See
The trunnion ring of the X-ray tube housing holds the X-ray tube housing assembly in two
drawing Z-11
clamp rings. It provides that the X-ray tube housing assembly can rotate around its longitu-
dinal axis. This movement can be blocked by tightening the two clamp rings.
The adjustment of the central beam in longitudinal and transverse direction with respect to
the fixing of the X-ray tube housing assembly must be carried out on the system.
14
Figure 4: Removement of the alignment ring
(05.0)
2004 Philips Medical Systems
ALL RIGHTS RESERVED
SRO 2550 ROT 350
1
2
CMT_012
SRO 2550 ROT 350
SRO_2550_ROT_350_Manual_English.fm

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Rot 350

Tabla de contenido