Disinfection
The disinfection method used must conform to valid legal requirements and guidelines for
disinfection and protection against explosion.
WARNING
Switch OFF the mains voltage before disinfection of the X-ray equipment.
If you use disinfectants which create explosive gas mixtures, these must
first have dispelled before restarting the X-ray equipment.
Wipe disinfection
Disinfect all parts of the X-ray equipment, including accessories and connecting cables, by
wiping only with a soft cloth.
Spray disinfection
Spray disinfection is not recommended because disinfectant could get inside the X-ray
equipment.
Disinfection by atomizing
Cover the X-ray equipment carefully with a sheeting when it has cooled down. After the
disinfection mist has subsided, you may remove the sheeting and disinfect the X-ray
equipment by wiping with a soft cloth.