Maintenance by the user

Visual check
Check the X-ray equipment for apparent defects:
Table 3: Defect check
Error
Check the X-ray equipment for error messages:
messages
Table 4: Error message check
In case of other error messages concerning the X-ray tube housing assembly call your lo-
cal service.
SRO 2550 ROT 350
SRO_2550_ROT_350_Manual_English.fm
Interval
As per relevant national and
international standards and
laws or the local regula-
tions.
Daily
Weekly
Weekly
Type
Meaning
Warning
The X-ray tube housing assembly is
thermally overloaded and cannot oper-
ate further without a rest period.
2004 Philips Medical Systems
ALL RIGHTS RESERVED
Scope of work
Stability test
Damaged parts,
labelling and warning
plates
All cables and
connections
(damage, breaks)
Oil leaks and unusual
noises
(05.0)
Method
As per relevant national and
international standards and
laws or the local regula-
tions.
Inspection
Inspection
Inspection
Action
Maintain rest intervals and
wait for enabling at generator
desk.