FUNCTIONAl ANOMAlY
The cycle does not start and
all control panel keys flash.
The luminous STOP key
flashes accompanied by an
intermittent acoustic signal.
On switch-on, all luminous
indicators on the control
panel flash for 3 seconds and
an acoustic signal is heard
(long beep).
The cycle starts (keys flash)
but the pump does not
start.
The machine does not switch
on.
64
TROUBlESHOOTING GUIDE
POSSIBlE CAUSES
Overheating, damage to the
electronic circuit.
Insufficient vacuum in the cham-
ber due to one of the following
reasons:
• lid not pressed on cycle start-
up.
• lid gasket worn or incorrectly
positioned.
• leak from vacuum circuit.
• Faulty vacuum sensor.
• Clogged vacuum pump.
Warning that oil change is requi-
red as 2500 functioning cycles
have been performed.
Pump circuit breaker protec-
tion intervention due to over-
heating.
No energy.
Instructions for Use
SOlUTION
• Switch the machine off and
wait for about 15 minutes for
cooling and automatic restore
of the circuit breaker protec-
tions.
If the problem persists when
switched back on, call the af-
ter-sales technical assistance
to check the machine.
• Press the lid upon cycle start-
up.
• Check and eventually replace
the lid gasket.
• Call the after-sales technical
assistance to check the air
circuit.
• Call the after-sales technical
assistance to check the vacu-
um sensor.
• Call the after-sales technical
assistance to check the vacu-
um pump.
• Intervention of a technician to
replace the pump oil.
• Switch the appliance off and
wait about 15 minutes before
re-applying voltage.
• Check the connection to the
electric line.
FVCD 270