7.1 Technical Data
Type:
Bores:
Display:
Operations:
Temperatures:
Temperature range:
Temperature stability:
Warm up time:
Heating times:
Time range:
Safety:
Interfaces:
Internal quality control
(IQC):
Languages:
Update:
Operating range:
Power supply:
Power consumptions:
Dimensions:
Weight:
Warranty:
CE certified:
Declaration of
conformity:
Overvoltage category:
Degree of pollution:
Manual heating block NANOCOLOR
38
Heating block for chemical‑analytical digestion
2 × 12 with 16 mm OD (VARIO 4)
8 with 16 mm OD + 2 with 22 mm OD (VARIO C2 M)
12 with 16 mm OD (VARIO C2)
Backlit, multicolor LCD touchscreen
Icon‑based menu guidance via touch screen
6 preprogrammed temperatures 70 °C / 100 °C / 120 °C / 148 °C / 150 °C / 160 °C, 6
free memory locations for individual temperature settings
40 °C–160 °C (1 °C increments)
± 1 °C (according to DIN, EN, ISO and EPA methods)
From 20 °C to 160 °C within 10 min (at 20 °C)
5 preprogrammed heating times, 10 min / 30 min / 60 min / 120 min / cont.
7 free memory locations for individual heating times
0 h:01 min–9 h:59 min (1 min increments)
Replaceable safety covers as contact protection, lockable protective lids, overheating
protection
Bidirectional RS‑232, USB A (function), USB B (host)
With NANOCOLOR
T‑Set (REF 919917) and NANOCOLOR
®
919921)
optional fully automatic calibration and generation of a test certificate for device control
and monitoring
DE / EN / FR / ES / HU / PL / CZ / TR / DK
via USB‑Stick
10 °C – 40 °C; max. 80 % relative humidity (non‑condensing), up to 3000 m
110 V / 230 V, ~50 / 60 Hz
300 / 550 W (VARIO 4)
150 / 300 W (VARIO C2, VARIO C2 M)
290 mm × 287 mm × 146 mm (VARIO 4)
169 mm × 282 mm × 146 mm (VARIO C2, VARIO C2 M)
ca. 3.2 kg (VARIO 4)
ca. 2.0 kg (VARIO C2, VARIO C2 M)
2 Years
Yes
On request
II
2
VARIO C2 / VARIO 4 EN, 08.2021
®
USB T‑Set (REF
®