15. Data export
The data generated in the course of quality control can be output to a PC using the "Ka‑
lib‑Software" program supplied. Proceed as follows:
1. Connect the heating block to a computer using the USB‑A/B cable provided (see scope of
delivery) and check the connection (device manager).
2. Start the "Kalib‑Software" from the supplied DVD. The connected heating block is automat‑
ically searched and the data is read in the background.
3. Click on "Testing" in the software to generate a certificate of temperature verification. Click
on "Calibration" to generate a certificate of heating block adjustment.
4. Then disconnect the device from the computer.
16. Update
We recommend to make a data backup before the update (see chapter "Data export").
MACHEREY‑NAGEL offers updates for the heating blocks on its website. Download the folder
with the update files from the MACHEREY‑NAGEL website. Unzip the folder and follow the
instructions in the update guide.
17. Maintenance and cleaning of the device
17.1 Cleaning of the device
For all cleaning work, the unit must be switched off and disconnected from the power
supply.
Fire hazard: Do not use flammable cleaning agents or organic solvents to clean the
unit.
Danger of burns: Do not clean the device while it is still hot.
Risk of chemical burns: If a cuvette breaks, make sure that the liquid does not come
into contact with the skin. If necessary, use an exhaust system to remove chemical
vapors.
Dispose of chemicals, waste and used cuvettes according to national, regional and
local regulations.
Manual heating block NANOCOLOR
54
NOTICE
NOTICE
WARNING
WARNING
WARNING
WARNING
VARIO C2 / VARIO 4 EN, 08.2021
®