PJX Series Balance
Output format
7.2
The Result Data, and G/N/T data, is output in the following format.
1
Field:
Label
Space
Length:
Note:
1. The length of the label field is not fixed.
2. Each field is followed by a single delimiting space (ASCII 32).
3. The Weight field is 11 right justified characters. If the value is negative, the '–' character is located at the
immediate left of the most significant digit.
4. The Unit field contains the unit of measure abbreviation up to 5 characters, right justified.
5. The Stability field contains the "?" character if the weight reading is not stable. The Stability field and the
following Space field are omitted if the weight reading is stable.
6. The G/N field contains the net or gross indication. For net weights, the field contains "N". For gross
weights, the field contains "G".
7. The Termination Characters field contains CRLF, Four CRLF or Form Feed (ASCII 12), depending on
the LINE FEED menu setting.
8. When Numeric Only is set On, only the Weight Field is printed, left-aligned.
Printout Examples
7.3
Examples for each Application are shown with all items turned ON in the Print menu. The default values for
Header lines 1-3 are also shown.
WEIGHING
Header 1
Header 2
Header 3
07/21/2017
Balance ID: B234567890
Balance Name: PJX8201/E
User Name: OHAUS
Project Name: PJX
Weigh
1365.1
Gross:
Net:
1365.1 g NET
Tare:
2554.8 g T
Signature: ____________
Verified By: ____________
2
3
Weight
Space
1
11
1
17:41:44
g N
3919.9 g G
2
4
Unit
Space Stability
5
1
5
6
Space G/N
Space
1
1
3
PART COUNTING
Header 1
Header 2
Header 3
07/21/2017
17:41:44
Balance ID: B234567890
Balance Name: PJX8201/E
User Name: OHAUS
Project Name: PJX
Count
Quantity:
201 PCS N
Gross:
300.1
g G
Net:
200.3
g N
Tare:
99.8
g T
APW: 1.00 g
Sample Size: 100 PCS
Signature: ____________
Verified By: ____________
EN-35
Term.
Characters
7
0
8