SECTION 9 – EMC COMPLIANCE
To maintain basic safety and the essential performance of the device during the expected service life, it is necessary to install and start it
according to the EMC information provided in the table. Keep the device away from HF surgical equipment and RF shielded rooms with
instruments such as magnetic resonance imaging, where the intensity of the EM disturbances is high. The portable RF communication devices
may affect the operation of the ES500 including the essential performance due to electromagnetic disturbances. Therefore, these devices
should be kept as far away as possible from the ES500 and its accessories in order to avoid such effects.
Emissions Test
RF Emission CISPR11
Immunity Test
Electrostatic Discharge
(ESD) IEC 61000-4-2
Radiated RF
IEC 61000-4-3
Conducted RF Immunity
IEC 61000-4-6
Power Frequency magnetic field
IEC 61000-4-8
Immunity to proximity fields from RF wireless communication equipment 60601-1-2 (Clause 8.10)
Test Level and Compliance Level are the same below:
Frequency Band
Test Frequency
(MHz)
380-390
385
430-470
450
704-787
710
745
780
800-960
810
870
930
1700-1990
1720
1845
1970
2400-2570
2450
Compliance
Group 1, Class B
IEC 60601 Test Level
±8KV - Contact
±2KV, ±4kV, ±8KV, ± 15KV - Air
3 V/m
80MHz-2.7GHz
3Vrms
0.15MHz – 80Mhz
6Vrms ISM bands
80% AM at 1kHz
30 A/m
Modulation
(MHz)
PM 18kHz
FM ±5kHz deviation
1kHz sine
PM 217kHz
PM 18kHz
PM 217kHz
PM 217kHz
Compliance Level
±8KV - Contact
±2KV, ±4kV, ±8KV, ± 15KV - Air
3 V/m
80MHz-2.7GHz
3Vrms
0.15MHz – 80Mhz
6Vrms ISM bands
80% AM at 1kHz
30 A/m
Maximum Power
(W)
1.8
0.3
2.0
0.3
0.2
0.3
2.0
0.3
2.0
0.3
2.0
0.3
Distance
Immunity Test
and Compliance
(m)
Level (V/m)
27
28
9
28
28
28
13