Descargar Imprimir esta página

Minebea Intec Midrics MIS1 Serie Instalación E Indicaciones De Seguridad página 10

Ocultar thumbs Ver también para Midrics MIS1 Serie:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 44
1. Power connection
Explosion-safe area
2. Data transfer
Explosion-safe area
PC, printer,
PLC, etc.
This equipment may also be
installed in zone 2 or zone 22
if they are suitable for
Category 3 as per the ATEX
Directive.
Written by
Reviewed by
Released by
10
Installation Instructions and Safety Information
Explosion-risk area zone 2 or zone 22
Gas:
Dust:
T
AMB
Use explosion-protected
Connector of
equipment
supplied
Protect the connector of the
equipment supplied from
being disconnected. Secure
the emergency stop.
Explosion-risk area zone 2 or zone 22
Gas:
Dust:
T
AMB
For data transfer only and, if required, supply for the
connected equipment (e.g. printer), there is no supply
voltage going from the connected equipment to the
indicator/complete scale
Date
Name
Material
19.02.2021
T. Hiller
22.02.2021
R. Koch
Revision
03
22.02.2021
T. Hiller
IIC, T4
IIIC, T80°C
for indicators
IIIC, T100°C
for complete scales
:
-10°C to +40°C
connector
6)
Potential equalization
IIC, T4
IIIC, T80°C
for indicators
IIIC, T100°C
for complete scales
:
-10°C to +40°C
6)
Potential equalization
---
Title
Alteration
Drawing number
543604
Alternative
connections
Indicator MIS.
or
complete scale
MW...-......-......
each + option Y2
Indicator MIS.
or
complete scale
MW...-......-......
each + option Y2
Scale
---
Sheet
Safety instructions
1
of
36418-740-16 (EN)
4

Publicidad

loading