Visual inspection and functional test
- In accordance with the IEC 62353-VDE 0751-1, these inspections/tests must only be per-
formed by qualified persons who have received adequate training and have access to the
required equipment.
Electrical measurement
- Electrical measurements according to IEC 62353-VDE 0751-1 may also be performed by spe-
cially trained and technically authorized persons (e.g. in-house technicians) with additional
medical and bed-specific knowledge if suitable measurement equipment is available.
- Only a qualified electrician with additional medical and bed-specific knowledge may evaluate
and document the test results.
IEC 62353-VDE 0751-1 a short checklist
Visual inspection
-
Mechanical components: (undercarriage, side
guards, wooden frame, patient surface frame,
lifting pole socket)
- Electrical components: power and handset
cable, handset
- Leakage current test (difference Δ I without load)
- Protective conductor testing (protection class 1)
9.1 Spare parts
Danger!
There is a highly increased risk of electric shock or fire when using incompatible
electric components. Only replace electrical components with original spare parts.
Only wissner-bosserhoff GmbH original spare parts may be used. Contact our customer ser-
vice, sales and technical consultation department for information concerning spare part deliv-
eries. On request, please provide the details on the nameplate and the order number.
Our qualified technical customer service will be happy to help:
Phone
+49 2377 784-456
Fax
+49 2377 784-150
Functional test:
- Mechanical components: castors, emergency
lowering of the backrest, side guards, calf rest,
motor bolts, screws
- Electrical components: Handset
Documentation
:
- Test protocol, non-conformity report, inventory list,
instructions for use