Alarm is triggered without any
apparent reason
Weight is not displayed correctly
or values of the weight history
differ too much
Display shows „hi" or „lo"
Underbed light is not working as
it should
Handset does not work
Mobilift
does not work
®
CareBoard
does not react
®
properly
The power box emits an
acoustic signal
Save & Send button lights up
red
Nursecall system adapter is
defective
Cable is defective
Sensors are defective
Missing magnets
Weight was added/removed
without freezing it
Weight was set to zero acci-
dentally
Wrong measuring
Zero point set incorrectly
Wrong standard saved (patient
is in the bed when turning on
automatic functions)
Handset locked, not correctly
plugged in or faulty
GO button is not pressed
Handset is locked
Mobilift is defective (cable
break)
You are pressing the touch
screen too hard or the area
being pressed is too large
CareBoard
not correctly
®
connected / faulty
Fault affecting the server
connection
Contact in-house technician,
otherwise contact the WiBo
customer service
Ignore extremely differing/
wrong values
Set the display to zero when the
resident is not in bed
Switching underbed light into
automatic mode when the
resident is in bed
Unlock handset, check connec-
tion. Otherwise contact the
customer service
Press the GO button (handset or
CareBoard
)
®
Unlock handset
Contact the WiBo customer
service
When operating the board, plea-
se make sure that you press the
center of the button and do not
push too hard
Check the plug connection or
call WiBo service
Check network cable, reset the
CMC box (see chapter 5.2) – or
contact WiBo service