P P r r e e v v e e n n t t i i v v e e m m a a i i n n t t e e n n a a n n c c e e
Remove product from service before you perform the preventive maintenance inspection. Check all items listed during
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks
more often based on your level of product usage. Service only by qualified personnel.
Clean and disinfect the exterior of the support surface before inspection, if applicable.
Inspect the following items:
All welds
All fasteners are secure
Brake mechanism works
Steer function works
Siderails raise, lower, and latch
Casters lock when you apply the brakes
Casters secure and swivel
Casters are free of wax or debris
Fowler backrest raises, lowers, and latches
Skins are not cracked
Trendelenburg/Reverse Trendelenburg raises and lowers from all locations
Foot section adjusts to each position
Enhanced clearance headpiece adjusts to each position, and locks and releases (optional)
Arm boards are intact and work
Arm board support levers are intact and work
IV pole is intact and raises and lowers to each position (optional)
Body restraints work (optional)
Accessories and mounting hardware are in good condition and work
No rips or cracks in mattress cover
V V e e l l c c r r o o ® ® on litter is in good condition; replace if necessary (0381-024-007)
Ground chain is intact
No leaks at hydraulic connections
Hydraulic jacks are holding
Lubricate where required
Product serial number:
Completed by:
Date:
5050-009-005 Rev C.0