Specifications
Display: 3-2/3–digit LCD (2499 maximum display)
Polarity: Automatic
Sampling Rate: 3 per second
Temperature Coefficient: Nominal 0.15 x (specified accuracy) per °C
below 18 °C or above 28 °C
Automatic Power-Off: After 30 minutes of inactivity.
To disable this feature, press Select while turning the meter on.
Noise Rejection*:
Normal Mode Rejection Ratio > 50 dB at 50 Hz and 60 Hz when measuring DCV
Common Mode Rejection Ratio > 60 dB from 0 Hz to 60 Hz when measuring ACV
Common Mode Rejection Ratio > 100 dB at 0 Hz, 50 Hz and 60 Hz when measuring DCV
Operating Conditions:
0 °C to 35 °C (32 °F to 95 °F), 0% to 80% relative humidity (non-condensing)
35 °C to 40 °C (95 °F to 104 °F), 0% to 70% relative humidity (non-condensing)
Altitude: 2000 m (6500') maximum
Indoor use only
Storage Conditions: –20 °C to 60 °C (–4 °F to 140 °F), 0% to 80% relative humidity
(non-condensing)
Remove battery.
Battery: Two 1.5-Volt batteries (NEDA 24A or IEC LR03)
Pollution Degree: 2
Overload Protections:
Volts: 780 V RMS, 1000 V peak
Millivolts: 600 VDC/VAC RMS
A: 12.5 A/500 V fuse, interrupting rating 20 kA, 1/4" x 1-1/4"
µA and mA: 0.63 A/500 V fuse, interrupting rating 200 kA, 1/4" x 1-1/4"
Other Functions: 600 VDC/VAC RMS
Overvoltage Categories:
+ Terminal: Category III, 600 VAC and 600 VDC; Category II, 750 VAC and 1000 VDC
µA and mA Terminal: Category III, 240 VAC; Category II, 150 VDC
A Terminal: Category III, 240 VAC; Category II, 150 VDC
18