Instructions for Use Vyntus WALK
Putting the Unit into Operation
Always consult the nameplate on the device/system for compliance of the unit's
own data with those of the local power supply system (mains voltage and mains
frequency) before actually connecting the unit.
Connect only if all data comply!
Inspect the mains connection cable, plug, and receptacle for visible damages
prior to establishing the connection. Damaged cables or plugs must be replaced
immediately.
Installation and assembly of the device must only be done in compliance with this
Instructions for Use.
After the first setup or if the setup has been changed (e.g. exchanging of
pneumotach), a function test (e.g. calibration) has to be performed.
Medical Supervision
A qualified physician must reassess all measurements. An interpretation by the
medical device is significant only when considered together with other clinical
findings.
The user is responsible for the change of predicted modules, authors or the entry
of new predicted value formulas.
Maintenance
Use only CareFusion approved accessories and spare parts for this medical device.
If applied parts (e.g. handle) have been exposed to extreme mechanical stress, a
function test (e.g. volume calibration) has to be performed. If function is lost, the
defective part is to be replaced. Damaged parts, e.g. frayed plugs, receptacles,
pressure regulators, a damaged handle, and defective cables should be replaced
immediately by an authorized specialist or engineers of the CareFusion Global
Customer Support. If it is opened without authorization the guarantee
entitlement expires. The CareFusion Global Customer Support is always at your
disposal with help and assistance in case of problems.
Keep disposables, accessories, packing material, cleaning and disinfection
substances out of the reach of children.
Swallowing packaging or pulling it over your head can result in death due
to suffocation.
Keep the packaging out of the reach of children or mentally restricted individuals
and dispose it according to local standards!
Recycling
Adhere to the national law of your country when disposing the medical device and
its accessories.