Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Icare HOME Instruction Manual
30 A/m
30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
of power frequency
magnetic field should
be used no closer than
15 cm (6 inches) to any
part of Icare HOME
(TA022), including
cables specified by
the manufacturer.
Otherwise, degradation
of the performance
could result.
The measurement
method of the Icare
HOME tonometer is
based on magnetic
induction and therefore
an external magnetic
field in line with the
probe may prevent the
measurement.
In such case the
tonometer will
continuously ask
to repeat the
measurement.
Situation can be solved
either by removing the
source of interference
from the vicinity of the
device or by performing
the measurement in
different location with
no such interference.
ENGLISH
WARNING: Sources