Quick Setup Guide TDL GPro 500
7 Error messages
Message
No sensor on channel 3
Signal Processing Failed
Laser Source Error
Bad Signal Quality
Flashcard Error
Pressure Input Error
Temperature Input Error
Configuration Mode
The GPro 500 error messages can be found in the M400 under the following path:
Menu ➝ Service ➝ Diagnostics ➝ TDL ➝ Messages
36
Comment
The M400 is unable to detect any of the ISM
If no sensor is found it will display the message NO SENSOR
DETECTED
Fitting of the line profiles failed.
The laser wavelength has shifted. Readjustment of the laser tempera-
ture necessary
Transmission lower than 5% threshold
Missing or bad calibration and / or database data
Pressure reading out of extended range: 0.1 bar(a) – 10 bar(a)
4 – 20 mA input error: 4 mA > P > 20 mA
Temperature reading out of extended range: – 20°C < T < 600°C
4 – 20 mA input error: 4 mA > T > 20 mA
Ethernet port in use: diagnostic or configuration in progress