2.
SAFETY INSTRUCTIONS
2.1
Intendend use
The luminaire Dmed® OPTICLUX Hand is an examination
luminaire. It is intended to illuminate the body of a patient
locally up to a distance of approx. 15 cm from the treating
surface in order to support the dermatological diagnosis and
treatment. The diagnosis can be interrupted at any time due
to a light failure without endangering the patient. The
luminaire is not intended for use in operating rooms.
In addition, the Dmed® OPTICLUX Hand can also be used in
the medical environment of a laboratory or during industrial
quality control.
2.2
User profiles
Medical Specialist
Are all persons who have completed medical training and
work in their trained professional field.
Cleaning specialist
Is familiar with national and workplace hygiene regulations.
Electrician
He is trained in electronics and electrical engineering and
knows the relevant standards and regulations.
Qualified specialist
Due to his technical training, knowledge and experience as
well as knowledge of the regulations, he is able to carry out
the assembly / disassembly.
2.3
Safety manual
►
Operation by health professionals
►
This manual is part of the product and must be stored
and made available to all future users.
►
All work on the luminaire (including repairs) must be
done only by qualified professionals.
►
The luminaire may not be altered or tampered with. Only
approved original parts may be used. Any use other than
that intended, using original parts, can change technical
parameters and cause life-threatening hazards.
►
Operation in combustible or explosion-prone areas is
prohibited. The luminaire's power supply is a potential
ignition source.
►
The luminaire must be used only in dry, dust-free rooms.
►
The luminaire should not be left on without supervision.
►
Do not use a damaged luminaire. Defective cord
locations also represent a potential hazard. Do not place
the cord near heat sources or sharp edges.
►
Eye damage: Never look directly into the light cone.
►
Replace damaged glass before operating the luminaire
again.
►
The luminaire must not be covered when operational.
►
The ventilation openings (if present) must always be kept
clear during operation!
►
The luminaire must not be operated near external heat
sources that exceed the luminaire's maximum ambient
temperature.
►
Always store the hand magnifying lamp in the protective
cover when not in use. Attention: (Fire and burn hazard)
►
The luminaire must not be used outside the specified
ambient conditions.
►
Do not use with medical devices that may react
sensitively to a light spectrum within the visible range
(such as pulsating light and/or light with a high
illumination intensity).
►
The luminaire may only be used for the intended use
described here.
►
The manufacturer cannot be held responsible for any
damages resulting from use deviating from the intended
use, or failure to observe the safety instructions and
warnings.
►
During an examination in which the eyes may come into
contact with UV light (Wood light version only), the user
must advise the patient to keep his or her eyes closed.
►
The luminaire is intended to last 10 years (not including
the battery).
When using several lights simultaneously, the total
illuminance Ee in the light field 1000W/m² must not be
exceeded during operation.
When using several lamps simultaneously, the maximum
UV illuminance Euv < 10W/m² must not be exceeded
during operation.
2.4
Warning levels
DANGER
Warning of hazards that can result in death or serious
injury if instructions are not followed.
WARNING
Warning of hazards that can result in injury if instructions
are not followed.
CAUTION
Warning of hazards that can cause material damage if
instructions are not followed.
ENG
15