4.
Allow the battery to charge until the indicator LED is solid green.
Remove the monitor from the cradle, and then press the Power button on the
5.
monitor.
Note: Do not attach a blade or baton at this time.
6.
In the upper right corner of the monitor screen, verify that the installed software
version is 1.3 or higher. If not, contact Verathon
update.
PROCEDURE 3. CONFIGURE USER SETTINGS
The User Settings Tool is a Java‑based tool and is available on the USB flash drive.
1.
Connect the USB flash drive to a USB port on a computer.
2.
Navigate to the USB flash drive, and then open the User Settings Tool.
Configure the settings as needed, and then click Save.
3.
4.
In the Save As dialog box, navigate to the USB flash drive, and then click Save.
5.
Ensure the monitor is powered off, and then insert the USB flash drive into the
micro‑USB port on the monitor.
On the monitor, press the Power button. The monitor powers on and the settings
6.
automatically update. The settings file is then automatically deleted to help prevent
accidentally overwriting the date and time settings.
PROCEDURE 4. INSERT THE VIDEO BATON INTO THE STAT (OPTIONAL)
If you are using a video baton and a GVL
connect the baton to the monitor.
1.
Open the GVL Stat pouch, but do not remove the Stat from the packaging.
2.
Ensure that the logo on the side of the baton and the logo on the side of the Stat
are aligned.
3.
Slide the video baton into the GVL Stat until it clicks into place. Do not remove the
Stat from the pouch until you are ready to begin the intubation. This ensures that
the Stat remains as clean as possible.
Note: Ensure that you do not insert the video baton backwards.
Correct
Customer Care for a software
®
Stat, attach the Stat to the baton before you
®
Incorrect
7