✓ The handpiece is recognized and the corresponding information displayed.
D Press the right footswitch pedal.
✓ The motor starts.
D Release the footswitch pedal.
✓ The motor stops immediately.
D Change the direction of rotation by pressing the mid button of the footswitch.
✓ The displayed direction of rotation changes.
For devices with motor socket M2:
D Connect a handpiece to the motor socket M2.
D Select the handpiece M2 on the touch screen.
✓ The handpiece is recognized and the corresponding information displayed.
D Press the right footswitch pedal.
✓ The motor starts.
D Release the footswitch pedal.
✓ The motor stops immediately.
D Change the direction of rotation by pressing the mid button of the footswitch.
✓ The displayed direction of rotation changes.
For devices providing a pump:
D Press the left footswitch pedal.
✓ The pump starts running.
D Change the flow of the pump via touch screen.
✓ The pump has to run regularly at all settings.
D Open the tensioning lever of the pump.
✓ The error message E16 Irrigation pump not closed is displayed.
D Close the tensioning lever of the pump again.
✓ The error message disappears.

4.2.4. System Information

D By pressing the wrench button switch to the Settings menu and choose System Information.
D Check the displayed system information.
✓ The serial number displayed on the touch screen has to comply with the one printed on the labelling
at the back of the housing cover.
D Note the displayed software versions.
D Return to the Settings menu by pressing the X button.

4.2.5. Test for proper operation ECO devices

D Connect the power cord and switch the device on.
✓ While switched on the power switch has to be illuminated.
✓ The error message E11 (no footswitch) is displayed.
4-4
05.13 / V1.0