Manual Cleaning And Disinfection - Erbe ERBECRYO 2 Indicacion De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 22
All manuals and user guides at all-guides.com
OBJ_DOKU-190376-005.fm Seite 13 Donnerstag, 26. November 2020 3:06 15
5.2

Manual cleaning and disinfection

Rinse the product with water immediately after use and wash it
with an alkaline cleaning solution.
Disinfect the product after cleaning. When doing so, always
check the instructions provided by the disinfectant manufacturer
to ensure that the agent is suitable and of the correct concentra-
tion.
Dry the product at a temperature of max. 75 °C.
5.3
Cleaning and disinfection by machine
If heavily soiled, this product can be cleaned using a chemical/
thermal process in the washing machine. This method can short-
en the product's life span. Do not exceed a temperature of
max. 75 °C! It is essential that the product is fixed securely in
place!
6
Disposal
Dispose of the product, packaging material and accessories (if
available) in accordance with the national guidelines and legis-
lation applicable in each case.
7
Symbols
Symbol
Explanation
Consult instruc-
tions for use
Catalogue number
Symbol
Explanation
Caution, consult
accompanying
documents
Batch code
13

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

20402-200Erbejet 2

Tabla de contenido