3. SPECIFICATIONS

PARAMETER
Light Output
Color Temperature
LED life
CRI
Brightness Control
Light guide adaptor
Input Power
Fuses
Operating Conditions
Storage Conditions
Dimensions
Weight

4. CERTIFICATIONS

PARAMETER
System Classification
Isolation
EMC Certifications
CE Marking
Degree of protection
against harmful ingress of
water
Degree of safety in the
presence of Flammable
Anesthetics
Mode of operation
LIT-258 Sunoptic Surgical
Rev. A
VALUE
2800 lm (min.)
6500 K (typ.)
30,000 hours (typ.)
70 (typ.)
PWM (Pulse-width Modulation) 0-100% dimming
Rotating turret with STORZ, ACMI, WOLF, and OLYMPUS
(if equipped)
100-240 VAC, 50/60 Hz
150 W (max.)
5x20 mm, 250V, 2A, Type F
32 to 86⁰F (0 to 30⁰C)
30 to 85% RH, non-condensing
700 to 1060 hPa
-4 to 140⁰F (-20 to 60⁰C)
0 to 95% RH, non-condensing
700 to 1060 hPa
11.2" x 4.5" x 13.2" (W x H x D)
28.5 x 11.4 x 33.6 cm (W x H x D)
7.55 lbs / 3.42 kg
FDA Class I, Device Listing D095692
EU Class I, Active device per Annex IX, rule 1
Type BF
CISPR 11 Class A, IEC 60601-1-2 4
CE Marking for MDD 93/42/EEC and AMD 2007/47/CE
IPX-0; no protection.
Equipment is NOT suitable for use in the presence of
flammable anesthetics.
Continuous
®