6. SETUP AND OPERATION

6.1. Device Setup

Place the Illuminator on a stable surface such as a cart, counter, stand, etc.
WARNING: DO NOT use the device in any environment with explosive or flammable gases.
WARNING: Avoid placing the device in an area where the illuminator may be splashed with
liquids.
WARNING: DO NOT obstruct the exhaust or cooling vents of the illuminator. User must
ensure that environment air temperatures surrounding the unit are within the allowable limits.
Make sure the power switch is in the OFF position.
Connect AC power cord to the appliance inlet located on the rear panel of the light source.
CAUTION: Use only power cords provided with the unit or cables approved for medical use.
WARNING: To prevent electric shock, connect power cords of peripheral equipment through
medical isolation transformers.
NOTE: When using a medical isolation transformer, ensure that the transformer has sufficient
power ratings. Ensure that the power cord is connected to mains power with a three-prong
plug.
Connect a fiberoptic light guide to the topmost position of the rotating turret, ensuring that
the fiberoptic end tip type matches the accepting port.
NOTE: The illuminator has a light guide sensing feature in which the main illumination LED
will not turn-on until a light guide is connected to the console.

6.2. Operation

After the power cable and light guide are properly connected, turn on the illuminator by
pressing the mains power switch located on the front panel. The power indicator light
within the switch should illuminate.
NOTE: Upon startup, the illuminator will default to minimum illumination brightness.
STANDBY Button
Activates or deactivates STANDBY mode. STANDBY mode allows the user to turn the
main LED ON or OFF while the unit is still energized.
Press and HOLD for at least one second for this operation.
LIT-258 Sunoptic Surgical
Rev. A
®