ELECTROMAGNETIC COMPATIBILITY
Guidance and manufacture's declaration - electromagnetic immunity
This monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of this monitor should assure that it is used in such an environment.
Immunity Test
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
26 • ENGLISH
For all ME EQUIPMENT and ME SYSTEMS
IEC 60601 Test Level
Table 2
Compliance Level
± 6 kV contact
± 8 kV air
3 A/m
Electromagnetic Environment Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.