Device Messages - ResMed AirCurve 10 ASV Guía De Usuario

Ocultar thumbs Ver también para AirCurve 10 ASV:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Device messages

Device message/possible cause
High leak detected, check your water tub, tub seal or side cover
Water tub may not be inserted properly.
Water tub seal may not be inserted properly.
High leak detected, connect your tubing
Air tubing may not be connected properly.
Mask may be fitted incorrectly.
Tubing blocked, check your tubing
Air tubing may be blocked.
SD card error, remove your card and press Start to begin therapy
SD card may not be inserted correctly.
Read only card, please remove, unlock and re-insert SD card
SD card switch may be in the lock (read-only) position.
All other error messages, for example, System fault, refer to user guide, Error 0XX
An unrecoverable error has occurred on the device.
Solution
Make sure the water tub is correctly inserted.
Open the water tub and make sure that the seal is correctly
inserted.
Make sure the air tubing is firmly connected at both ends.
Make sure your mask is fitted correctly. See your mask user
guide for fitting instructions or use the Mask Fit function to
check your mask fit and seal.
Check the air tubing and remove any blockages. Press the
dial to clear the message and then press Start/Stop to
restart the device.
Remove and reinsert the SD card.
Move the switch on the SD Card from the lock position
to the unlock position
Contact your care provider. Do not open the device.
and then re-insert it.
English
13

Publicidad

Tabla de contenido
loading

Tabla de contenido