Air tubing
Air tubing
ClimateLineAir
ClimateLineAir Oxy
SlimLine
Standard
Heated air tubing temperature cut-out: ≤ 106°F (≤ 41°C)
Notes:
• The manufacturer reserves the right to change these specifications without notice.
• The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should
not be fitted to the mask.
• Do not use electrically conductive or antistatic air tubing.
• The temperature and relative humidity settings displayed are not measured values.
Displayed values
Value
Pressure sensor at air outlet:
Mask pressure
Flow derived values:
Leak
Tidal volume
Respiratory rate
Minute ventilation
Value
Pressure measurement:
Mask pressure
2
Flow and flow derived values
Flow
Leak
2
Tidal volume
2,3
Respiratory rate
2,3
Minute ventilation
2,3
1
Results are expressed as STPD (Standard Temperature and Pressure, Dry).
2
Accuracy may be reduced by the presence of leaks, supplemental oxygen, tidal volumes <100 mL or minute ventilation <3 L/min.
3
Measurement accuracy verified as per ISO 10651-6:2009 for Home Care Ventilatory Support Devices (Figure 101 and Table 101) using
nominal ResMed mask vent flows.
Measurement system uncertainties
In accordance with ISO 80601-2-70:2015 the measurement uncertainty of the manufacturer's test equipment is:
For measures of flow
For measures of volume (< 100 mL)
For measures of volume (≥ 100 mL)
For measures of static pressure
For measures of dynamic pressure
For measures of time
Note: ISO 80601-2-70:2015 stated accuracies and test results provided in this manual for these items already include the
relevant measurement uncertainty from the table above.
Material
Flexible plastic and electrical components
Flexible plastic and electrical components
Flexible plastic
Flexible plastic
Range
4–25 cm H
0–120 L/min
0–4000 mL
0–50 BPM
0–30 L/min
Accuracy
±[0.5 cm H
:
1
±6 L/min or 10% of reading, whichever is greater, at 0 to 150 L/min positive flow
±12 L/min or 20% of reading, whichever is greater, 0 to 60 L/min
±20%
±1.0 BPM
±20%
O
2
1
O + 4% of measured value]
2
± 1.5 L/min or ± 2.7% of reading (whichever is greater)
± 5 mL or 6% of reading (whichever is greater)
± 20 mL or 3% of reading (whichever is greater)
± 0.15 cm H
O
2
± 0.27 cm H
O
2
± 10 ms
Length
Inner diameter
6'6" (2 m)
0.6" (15 mm)
6'4" (1.9 m)
0.75" (19 mm)
6' (1.8 m)
0.6" (15 mm)
6'6" (2 m)
0.75" (19 mm)
Display resolution
0.1 cm H
O
2
1 L/min
1 mL
1 BPM
0.1 L/min
English
19