Applied
8.0
9.0
Press.
(mm)
(mm)
1.0 ATM
7.23
8.35
1.5 ATM
2.0 ATM
7.39
8.47
2.5 ATM
3.0 ATM
7.50
8.61
3.5 ATM
4.0 ATM
7.66
8.74
5.0 ATM
7.76
8.81
6.0 ATM
7.87
8.86 10.00 12.06 14.17 15.00 16.55 18.73
7.0 ATM
8.08
8.91 10.14 12.31 14.40 15.41 16.86 19.09
8.0 ATM
8.14
8.95 10.30 12.49 14.63 15.78 17.17
9.0 ATM
8.18
9.00 10.44 12.66 14.82
10.0 ATM
8.23
9.09 10.56 12.83 15.06
11.0 ATM
8.28
9.14 10.66
12.0 ATM
8.35
9.24 10.77
13.0 ATM
8.42
9.31 10.87
14.0 ATM
8.49
9.37
15.0 ATM
8.58
The upper figures in bold face represent the balloon diameter @ Nominal Inflation Pressure.
The lower figures in bold face represent the balloon diameter @ Rated Burst Pressure.
FOR ALL NUMED CATHETERS AN INFLATION DEVICE WITH PRESSURE GAUGE SHOULD BE USED.
Z-MED II-X™ Balloon Sizing Chart
10.0
12.0
14.0
15.0
(mm)
(mm)
(mm)
(mm)
8.97 10.80 12.63 13.02 14.73 16.61 18.65 20.88 22.03 23.52 26.34 28.48
9.19 11.01 12.90 13.28 15.05 17.01 18.98 21.52 22.86 24.34 27.04 29.45
9.38 11.25 13.22 13.65 15.44 17.49 19.29 22.19 23.54 25.21 27.69 30.41
9.58 11.54 13.55 14.08 15.84 17.97 19.93 22.79 24.18 25.93
9.79 11.84 13.85 14.53 16.24 18.38 20.46
16.0
18.0
20.0
(mm)
(mm)
(mm)
(mm)
4
22.0
23.0
25.0
28.0
(mm)
(mm)
(mm)
26.74 28.95
27.39 29.96
27.97
30.0
(mm)