parameters. You can modify these settings afterwards, however,
only within a range that is not of critical importance to the prod-
uct.
After plugging in the product, "Bipolar resection" mode must be
displayed. Before activating or changing the operating parame-
ters, make sure that the instrument has been correctly detected.
6
Cleaning, Disinfection, Sterilization
6.1
Safety Instructions
May be cleaned in an ultrasonic bath.
On no account use sharp or abrasive objects for cleaning purpos-
es.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 29 psi (2 bar).
Maximum water pressure during rinsing: 29 psi (2 bar).
During machine cleaning and disinfection, do not exceed 203 °F
(95 °C).
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of Validated
Procedures".
During sterilization, do not exceed 280 °F (138 °C).
Do not sterilize in hot air oven.
6.2
Reprocessing Limitation
Frequent reprocessing can damage this product. Observe the
safety instructions for checking the product. With visible damage
or functional impairment, the product may no longer be used.