OBJ_DOKU-194554-005.fm Seite 35 Mittwoch, 1. April 2020 3:21 15
6.3
Disassembly
1. First, separate the cable from the instrument. Do not damage
the cable insulation while doing so.
6.4
Required Aids
Reprocessing step
Precleaning
Manual cleaning/
disinfection
6.5
Precleaning
Use tap water for precleaning, a non-fixing disinfectant if neces-
sary.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
2. Ensure that the golden contact surfaces on the plug are not
visibly contaminated.
Items Needed
Soft plastic brush/soft cloth
Soft plastic brush/soft disposable
cloth (sterile, lint-free)
Compressed air gun (2 bar maxi-
mum)
35