Descargar Imprimir esta página

KERR ENDODONTICS elements IC Manual De Instrucciones página 8

Ocultar thumbs Ver también para elements IC:

Publicidad

Cartridge Level Indicator
4
a. The amount of gutta-percha in the elements gutta-percha cartridge
can be checked by looking at the Gutta-Percha (GP) level indicator.
The five increments indicator (4-3-2-1-0) provides the user with an
estimate of the residual amount of gutta-percha in the cartridge:
Indicator
Gutta-Percha Level
4
3
2
1
0
Retracting the Plunger
5
a. To retract the plunger before it reaches
the end of its stroke, press the Reverse
button. This will cause the plunger to
fully retract to its "home" position.
Replacing the Elements Gutta-Percha Cartridge
6
a. To replace the elements gutta-percha cartridge before it is completely
empty, first ensure that the unit is fully heated, then press the Reverse
button to retract the plunger.
b. Remove the cartridge by turning the nut 90° counterclockwise (CCW),
the pull the cartridge out of the device and dispose of properly.
90°CCW
WARNING
Handle empty elements gutta-percha cartridges by the plastic locknut only.
Other portions could be hot and could result in a serious burn.
ADVANCED SETTINGS
Changing Temperature Presets
1
The device has two different preset temperature modes.
(See the table in Step 4 of the GETTING STARTED section)
a. Press the Mode button to toggle to the
temperature setting CUSTOM.
b. Hold down the Mode button for
approximately 2 seconds to enter
Temperature Control Mode.
c. Press the Mode button repeatedly to
change temperature.
d. Once desired temperature is reached, hold down the Mode button
(approx. 2 seconds) again or wait for 4 seconds (non-activity) to exit
Temperature Control Mode.
PRECAUTION (BACKFILL ONLY)
Keep temperature setting above 150°C. Extruding material at lower
temperatures will create more stress on the motor, and may shorten the
device's life.
Adjusting the Audible Signal Level
2
a. Change the audible signal level by simultaneously holding down the
Mode and On/Off Power buttons for approximately 2 seconds. This will
open Audible Signal Control Mode.
b. Press the Mode button repeatedly to change the audible signal level.
c. Once you have set your desired level, exit Audible Signal Control
Mode by repeating the first step (a) above or wait for 4 seconds
(non-activity).
Backfill - Hold 2 Seconds
077-0901_REVA_ELEMENTS_IC_EU_IFU_BOOK_P24.indb 5
Full
75%
50%
25%
Empty
Downpack - Hold 2 Seconds
Adjusting the Motor Speed
3
The motor of the Backfill device has two speed settings.
a. Press Reverse button for approximately 2
seconds to enter Speed Control Mode.
b. Press Reverse button repeatedly to
change the motor speed (X1 - Standard
speed, X2 - Double speed).
c. Once you have set your desired speed,
exit Speed Control Mode by holding down
the Reverse button (approx. 2 seconds) or
waiting for 4 seconds (non-activity).
MAINTENANCE
Charging the Device(s)
1
The devices should be placed onto the Charging Base during and
between uses.
Replacing the Battery
2
Contact Customer Care if the battery appears to be malfunctioning. To replace
battery:
a. Remove silicon ball on the battery cover.
b. Remove screw on battery cover and remove battery.
c. Install new battery in same orientation.
d. Dispose of used battery per local regulations.
e. Cover the screw with the silicon ball.
PRECAUTION
Only use Kerr Endodontics batteries designated for this system. Use of other
batteries may damage the device(s) or cause malfunction.
CLEANING, DISINFECTION AND STERILIZATION INSTRUCTIONS
elements™ IC obturation system, Cartridge Needle Bender and
Heat Pluggers should be inspected prior to cleaning for the appearance of
defects such as cracks, deformations, corrosion, which are indicators that
the instruments are not in conditions to be re-used with the required level
of confidence.
The use of automated cleaning devices or automated disinfecting devices is
not recommended for the cleaning of the above components.
Health Care facilities are responsible for making sure that the
sterilization equipment is calibrated according to the manufacturers manuals
and specifications. In addition, health care facilities are responsible for
training their staff on infection control, proper
sterilization and disinfection procedures.
NOTE: make sure that the sterilization pouches are suitable for steam
sterilization and comply with your national guidelines, standards and
requirements.
- ISO 11607
- For USA: Use FDA-cleared accessories
PRECAUTION
• The electronic components should not be autoclaved as it will damage
the circuitry.
• Do not spray the devices with any liquids as it may damage the circuitry.
• Do not immerse the gutta-percha cartridges in any liquid.
• Do not allow liquids to collect in the Charger Base or come in contact
with the connectors as it may damage the circuitry.
• Do not allow liquids to enter the openings for the heat plugger on the
Downpack device or the elements gutta-percha cartridge on the backfill
device as it may damage the circuitry.
Hold 2 Seconds
5
6/13/19 6:21 PM

Publicidad

loading

Este manual también es adecuado para:

077-0901