Flash-overs, activation in the trocar or without visual control of
the distal end of the instrument can cause unwanted thermal in-
juries to the patient. Therefore, check that the distal end of the
instrument protrudes at least 10 mm out of the trocar before
each activation.
7
Cleaning, disinfection, sterilization
7.1
Safety instructions
Not suitable for manual cleaning and disinfection.
Suitable for precleaning using an ultrasonic bath.
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
It is essential to follow the cleaning agent and disinfectant man-
ufacturers' instructions.
During sterilization, do not exceed 134 °C.
Do not sterilize in hot air.
7.2
Reprocessing limitation
Frequent reprocessing can damage this product. Observe the
safety instructions for checking the product. With visible damage
or functional impairment, the product may no longer be used.