8-1 Cleaning (Handpiece)
1) Remove blood and other debris from the surface of the handpiece with clean water.
2) Wipe off any remaining moisture on the handpiece.
3) Wipe clean with alcohol-immersed cotton swab or cloth.
This icon denotes that the product can be washed via Thermo Disinfector.
Refer to the Thermo-Disinfector manual.
CAUTION
• After washing with Thermo-Disinfector and prior to lubrication, dry the product until all internal moisture is thoroughly removed. Thermo-
Disinfector moisture remaining inside the product could reduce the effect of lubrication and could cause corrosion inside of the product.
• To clean the product never use any solvent such as benzine or thinner.
• When cleaning debris off the handpiece surface, make sure no water enters from the rear side of the handpiece as entry of water may cause
handpiece malfunction.
• Do not rotate the handpiece in water as such operation may cause handpiece failure.
10