9.
Specifi cations
9.1
Equipment Classifi cation
Type of protection against electric shock.
Degree of protection against electric shock
Mode of operation.
Degree of protection against harmful ingress of
particles and/or water.
Degree of safety of application in the presence of
a fl ammable anaesthetic
*For home use, this can be upgraded to IPx2 when using the protective pouch (ACC-OBS-080).
9.2
FHR Performance
Standard Mode
Smoothed Mode
Manual Mode
9.3
General
Charger- 'R' models only
(Part No. 772559)
Max. Audio Output
Auto shut-off
Headphone output
USB Port
Real Time clock battery
Battery Type
Battery Life
Size
Internally powered equipment
Type BF - equipment with an applied part
Continuous
Main Unit: IP20*,
SR2/SR3 probes : IPX7
Protective Pouch : IP22
Equipment not suitable for use in the presence of a
FLAMMABLE ANAESTHETIC MIXTURE WITH AIR,
OXYGEN OR NITROUS OXIDE
Range: 60 - 210 bpm
Averaging: 4 beats
Range: 60 - 210 bpm
Averaging: 8 beats
Range: 60 - 210 bpm
Averaging: 10 beats
Protection :
Input Voltage :
Output Voltage :
Input Frequency :
Standby power consumption: 230V AC ≤ 0.1W
500 mW rms typical (loudspeaker)
3 minutes
Max. output Power:
25 mW rms (32Ω)
Connector:
3.5 mm stereo jack socket
Micro USB
SD Card Slot
RENATA CR1025, 3V Lithium
LR6 (Alkaline cells 1.5V) or HR6 (NiMH rechargeable 1.2V)
Typically, 500 x 1 minute examinations
Note: Battery life is typically 2 years or 500 charge/discharge cycles
140 x 33 x 75 mm
Weight
Resolution: 1 bpm
Accuracy: ± 3 bpm
Range
Resolution: 1 bpm
Accuracy: ± 3 bpm
Resolution: 1 bpm
Accuracy: ± 3 bpm
Class II
100-240 V AC ±10%
5VDC ± 5%
50 - 60Hz
Micro SD
280 g
EN