6.5
Battery Charging
•
Only HR6 (NiMH) batteries can be charged. Check the battery type before connecting the charger.
•
Do not attempt to recharge normal alkaline batteries. They may leak, cause a fi re or even explode.
•
Only use the charger and USB lead supplied by Huntleigh.
•
Switch off the Doppler before charging.
•
Do not use the Doppler on patients when the charger is connected.
If the battery level is low, a
During charging, a
When fully charged, the symbol will change to
6.6
Data transfer to a PC
The Doppler does not contain patient information however, stored traces and data can be transferred to a PC
running Huntleigh reporting software, via the same USB socket as for battery charging. Consult your local sales
representative for details.
Note: Please observe the safety warnings in the Huntleigh software IFU.
6.7
After Use
Press and hold the On/Off button for one second to switch the unit off .
Refer to the cleaning section before storing or using the unit on another patient.
7.
Care and Cleaning
7.1
General Care
The Doppler contains delicate components, for example the probe tip, which should be handled and treated
with care. Periodically, and whenever the integrity of the system is in doubt, carry out a check of all functions
as described in the relevant section of this IFU. If there are any defects, contact Huntleigh or your distributor for
repair or to order a replacement.
Cautions
•
Check with your facility's local infection control policy and medical equipment cleaning procedures.
•
Observe warnings and guidance on cleaning fl uid labelling regarding use and personal protective
equipment (PPE).
•
If detergent or disinfectant wipes are used ensure that excess solution is squeezed from the wipe
prior to use.
•
Always switch off the Doppler and disconnect from the AC supply before cleaning and disinfecting.
•
Always wipe off disinfectant using a cloth dampened with clean water.
•
Do not allow any fl uid to enter the products and do not immerse in any solution.
•
Do not use abrasive cloths or cleaners.
•
Do not use automatic washers or autoclaves.
•
Do not use Phenolic detergent based disinfectants, solutions containing cationic surfactants,
ammonia based compounds or perfumes and antiseptic solutions.
•
Insert the supplied charger lead into the USB socket at the top of the
Doppler.
•
Connect the charger and switch on mains power.
Charging will take approximately 5 - 6 hours, depending on battery
status.
symbol will appear on the Status bar.
symbol will be displayed on the screen.
.