Guidance and manufacturer's declaration
The CERTAS Plus Electronic Tool Kit is intended for use in the
electromagnetic environment specified below. The customer or the user
of the Tool Kit should ensure that it is used in such an environment.
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Table 2
– electromagnetic immunity
IEC 60601
test level
± 8 kV contact
± 15 kV air
Not applicable
Not applicable
Not applicable
30 A/m
15
Compliance level
± 8 kV contact
± 15 kV air
Not applicable
Not applicable
Not applicable
30 A/m
Electromanetic
environment –
guidance
Floors should be
wood, concrete
or ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30 %.
Not applicable
Not applicable
Not applicable
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.