On starting, the Hi_Cal will automatically check the probe measuring detection
sensor (Init), then move the probe up and down for setting the system refe-
rence.
In case of a problem during initialization, please proceed with following
controls:
•
Is the probe free to move ?
•
Is the measuring force correctly set ? (see chapter « Selector adjustement »)
•
In the event of uncontrolled movement (without using the UP/DOWN button)
of the carriage, switch the instrument off by pressing >5s on ON | OFF and
switch it on again by pressing >0.5s on ON | OFF.
Intuitive and interactive keypad
The Hi_Cal display electronic has been designed to make its use very easy,
following the concept of «start and go». All the main functions are accessible
by direct use of the corresponding button.
MIN
Setup
MAX
Quit
DEL
SET
Clear
Keypad direct function
ON
OFF
|
SET : Zero (long press)
SET /CLEAR
CLEAR : Initialisation of Min | Max during Min | Max measuring mode
Displays and activates the constant entry mode
Exchanges the reference Ref I ↔ Ref II
REF
Changes the active mode SURFACE ↔ DIAMETRE
Selects MIN | MAX | DELTA measurement modes
MIN / MAX / DEL
8
The Setup key enters a menu sequence which enable
access to parameters and extended functions
REF
During the menu sequence:
I / II
•
UP and DOWN keys select the next or previous
menu
•
LEFT and RIGHT keys select settings within
the active menu.
Stand-by
|
preset recall (short press)
|