Cleaning, Disinfection and Sterilization
Operation
7
Inspection
8
Packaging
30/645
Operating mode
Inspect devices and
sort out those with
defects.
Assemble the devices
(stops).
Place the devices in a
support or container
to avoid any contact
between instruments
or posts and pack the
devices in
"Sterilization
pouches".
END-IFU-PROPEX-PIXI-V15-WEB-EUR-Multilingual-2021-10
Warning
Dirty instruments must be
cleaned and disinfected
again.
Discard instruments which
show any deformations
(bent, twisted), damages
(broken, corroded) or defects
(loss of colour coding or
marking) affecting the
resistance, the safety or the
performance of the
instrument.
Do not use instrument
lubricants.
Avoid any contact between
instruments during
sterilization. Use supports or
containers.
Check the validity period of
the pouch given by the
manufacturer to determine
the shelf life.
Use packaging which are
resistant up to a temperature
of 141°C (286°F) and in
accordance with
EN ISO 11607.