PROBLEM
LCD shows "Er 5"
LCD shows "Er 6"
LCD shows "Er 7"
LCD shows "Er 8"
LCD shows "Er 9"
LCD shows "Er A"
LCD shows "Hi"
LCD shows "Lo"
TROUBLESHOOTING 2.2
POSSIBLE CAUSE
Cuff pressure above 300mmHg
More than 3 minutes with cuff
pressure above 15 mmHg
EEPROM accessing error
Device parameter checking error
MCU self-verify error
Pressure sensor parameter error
Systolic or diastolic measurement is
above specified range. Systolic: 60-
260mmHg Diastolic: 40-199mmHg
Systolic or diastolic measurement is
below specified range. Systolic: 60-
260mmHg Diastolic: 40-199mmHg
.
-25-
SOLUTION
Measure again after
five minutes. If the
monitor is still abnormal,
please contact the local
distributor or the factory
Make sure the cuff is
applied correctly and try
again; don't move while
taking measurement