Settings
12.1.3 Result Memory
12.1.4 Remember information
12.1.5 Dilution Format
12.1.6 Lamp Settings
32
Spectrophotometer NANOCOLOR
If the Reaction Time is activated, it can be skipped within a mea-
surement by pressing
By clicking on the Memory entry you can view the memory set-
tings. In addition to an overview of the memory usage, the mea-
surement result memory can be switched on or off here. By setting
the slider, the memory is activated or deactivated accordingly.
Clicking on the Remember Informations entry opens a window for
setting this function. If the respective option is activated, a value
entered once (e.g. for sample location) is taken over for each fur-
ther measurement until it is deselected or changed again. The
activity of this function can be recognized by the
status bar.
By setting the Menu entries setting, information such as the last
used scan area or LOT number is automatically called up the next
time it is used.
Clicking the Dilution format entry opens a window for setting the
dilution format. The formats "1 + X" and "1 : X" can be selected.
When activated, the selected dilution is displayed in this format
and taken into account in all menus with input of a dilution.
Incorrect measurement values due to not performed lamp
calibration. The halogen lamp must be recalibrated after a lamp
change.
Clicking the Lamp Settings opens a window for the lamp settings.
In this menu the working time of the halogen lamp is displayed. In
case of a lamp change the working time can be reset by clicking
the icon
.
When changing the halogen lamp it is mandatory to calibrate
the new lamp. After changing the lamp, press the icon
confirm the security question. The lamp calibration is performed
automatically.
Advance 1.0, 11.2020
®
.
NOTICE
icon in the
and