Results
14. Results
If the maximum memory usage is exceeded, the oldest mea-
surement results are overwritten. Before the memory is full, a
warning appears in the start screen with the reminder to save
the results.
The measurement data memory can store up to 1000 measure-
ment data, 1000 IQC results and 100 scans. All entered sample
information is stored in addition to the measurement result. The
sample information is linked to the measurement result during the
storage process in a way that is tamper-proof (GLP-compliant).
The results are recalled by selecting
The IQC memory saves all measurement results that are gener-
ated via the IQC menu or the IQC selection in the results window.
These are results of the internal quality control, which are stored
separately from the general Measurement Results.
The IQC memory can be called up via the
(see 13.7 IQC, page 37) or by selecting the IQC Measurement
Results tab in the measurement data memory.
Besides the type of IQC measurement, the associated method is
displayed. Each measurement is evaluated as Passed (green) or
Failed (red) .
You can scroll through the individual entries of the measurement
data or IQC-memory by scrolling the display. Detailed information
on a measurement can be obtained by selecting an individual
entry in the memory. In addition to the result, all entered sample
information is displayed here. Up to two results with detailed in-
formation can be opened simultaneously. In the case of a scan or
kinetics measurement, the recorded spectrum or the evaluation
graph of the kinetics is displayed in the detailed view.
Pressing
will display the options to export, delete or print the
results. In each case, a dialog box appears asking whether the
selected option should be applied to the selection or to the entire
data set.
Spectrophotometer NANOCOLOR
NOTICE
.
icon in the IQC menu
Advance 1.0, 11.2020
®