INSTRUCTIONS FOR USE
4.4 Cleaning and Disinfection Instruction:
WARNING:
Do not use bleach or products that contain bleach to clean the device. Injury or
equipment damage can occur.
After each use, clean the device with alcohol-based wipes.
Do not put the device into water. Equipment damage can occur.
Use a cloth and a quaternary ammonium disinfecting/cleaning solution to clean and
disinfect the device.
Read and follow the manufacturer's recommendation for low-level disinfection.
Read and follow the cleaning product's instructions. Use caution in areas where liquid
can get into the mechanism.
Wipe the device with a clean, dry cloth.
Make sure that the device is dry before you store it or use it again.
CAUTION: DO NOT IMMERSE PADS IN ANY LIQUID
CAUTION: DO NOT USE BLEACH OR PHENOLICS ON PADS
5. List of Applicable Standards:
Sl. no
Standards
1.
EN 62366-1
2.
EN ISO 14971
3.
EN 1041
4.
EN ISO 15223-1
5.
EN ISO 10993-1
6.
IEC 60601-2-46
Document Number: 80028153
Version C
Medical devices - Part 1: Application of usability
engineering to medical devices
Medical devices- Application of risk management to
medical devices.
Information supplied by the manufacturer of medical
devices
Medical devices - Symbols to be used with medical
device labels, labelling and information to be supplied
- Part 1: General requirements
Biological evaluation of medical devices - Part 1:
Evaluation and testing within a risk management
process
Medical electrical equipment - Part 2-46: Particular
requirements for the basic safety and essential
performance of operating tables
Page 15
Description
Issue Date: 30 MAR 2020
Ref Blank Template: 80025117 Ver. F