7. Edit
Press ▼or▲ to other pages.
Plate: working plate. plate No.: 1-7
Temp.: the working temp. The maximum setting is 130.
Mixing (min): the time for mixing. The maximum setting is 999.9.
Mixing (rpm): the speed for mixing.
The setting range is 500-3600.
Collect (sec): the time for collect. The maximum setting is 999.
Vapor (min): the time for vapor. The maximum setting is 999.9.
Pause: on/off to pause program
Spin tips plate is fixed on
plate 8. Make sure that the
plate of functional name is
assigned to TIP.
Plate 1 Name must be LB