The AccelStim device can be sensitive to electrostatic discharges with a value
> = 4kV. In the presence of such discharges the treatment in progress could be
paused. In this case, the user must press the button below the display to restart
the treatment.
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The AccelStim device is intended for use in the electromagnetic environment specified
below. The customer or the user should assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The AccelStim device is intended for use in the electromagnetic environment specified
below. The customer or the user should assure that it is used in such an environment.
Immunity Test
Electrostatic
1
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
¹ The AccelStim device can be sensitive to electrostatic discharges with a value > = 4kV. In
the presence of such discharges the treatment in progress could be paused. In this case, the
user must press the button below the display to restart the treatment.
Compliance
Group 1
Class B
Class A
Complies
IEC 60601
Test Level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
± 2 kV for
± 2 kV for
power supply
power supply
lines
lines
± 1 kV for
± 1 kV for
input/output
input/output
lines
lines
Electromagnetic Environment-
The AccelStim device uses RF energy
only for its internal function. Therefore,
its RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.
The AccelStim device is suitable for use
in all establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Compliance
Level
Environment - Guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be at
least 30%.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Guidance
Electromagnetic
26