Instructions for Use
Set-up
For the safety of the operator, the IPCMini (version 100-240Vac) controller
WARNING
must be powered with a 3-wire power cable connected to a plug approved at
international level. Use this cable and plug together with an adequately
earthed socket so as to prevent electrical shocks and satisfy the
requirements of EC norms. The high voltages that develop in the controller
can provoke serious injuries or death. After switching off the unit, residual
energy is present inside the controller for some time. Wait approx. 1 minute
to be sure that the residual energy has been dissipated.
The controller must be installed in a way that allows an easy interruption of
WARNING
the line voltage.
If this equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
For 100-240 Vac version.
WARNING
Power supply cord: the correct cable for electrical wiring is a three wires
cable (L+N+PE).
The wire section has to be at least AWG18, 0.83 mm
The power supply used for the DC 24 V Version shall provide separation
from mains voltages by reinforced or double insulation.
Power supply cord: the correct cable for electrical wiring is three wires
cable (L+N+PE). The Wire section has to be at least AWG18, 0.83 mm
PIN 1= Positive connector
PIN 2= Negative Connector
PIN 3= Protective Earth
Input Voltage 24Vdc ±10%
Agilent IPCMini Ion Pump Controller User Manual
2
. For 24 Vdc version.
2
155