Technical Information
Advanced Settings
Analog output
Agilent IPCMini Ion Pump Controller User Manual
Select the values for:
Protect value: see dedicated paragraph.
Power limit: output power to the pump can be set at lower value
than the maximum of 40 Watts.
iSTEP: see dedicated paragraph
For RS232 or RS485 selection.
When RS485 is selected, addresses
from 0 to 31 can be set.
When LAN (Ethernet) is selected
IP address and the Subnet mask can
be set.
Analog Output can be set according to:
- Current: analog output proportional to the
current logarithm.
- Pressure: analog output proportional to
the pressure logarithm.
- Voltage: analog output linearly proportional to
the Voltage.
- MiniVac: it duplicates the MiniVac output
signal values (linearly proportional to the
current).
Minimum and maximum range values can
be selected.
For more details, see paragraphs
"Analog output signal from IPCMini" and
"Compatibility with old Minivac controller".