Technical Information
Set- point
The controller has 1 relays and they can be configured as set-points.
The microprocessor begins to drive the relay "Mask time" seconds after reaching
the target voltage. "Mask time" is a value that can be set in the "Set-point" menu.
This feature avoids the incorrect activation due to voltage ramping.
The set-point threshold has a configurable hysteresis (in following example of 5%).
Figure 49
Autostart
This function allows to restore the high voltage output when the main power supply
is restored after a fail.
If the Autostart is enabled:
When the main power fails and the hv channel is in on status, after the main is
restored, the hv turns on automatically;
When the main power fails but the hv channel is in off status, after the main is
resotred, the hv channel remains disabled.
If the Autostart is disabled:
After the power is restored after a fail, the hv remains disabled.
Agilent IPCMini Ion Pump Controller User Manual