Instructions for Use
If an HV cable without interlock is to be used, the accessories bag contains a
NOTE
specific kit of preassembled interlock lead. However, with this type of use, the
safety capability furnished using the interlock is lost.
Refer to the figure below for correct assembly of the interlock cable furnished in the
accessories bag in the case in which an HV cable with interlock is not used.
Figure 39
The cable interlock closes on pump ground. The high voltage is disabled if the
NOTE
connection is interrupted.
The cable interlock closes on pump ground. The high voltage is disabled if the
connection is interrupted. To connect the controller to the pump, use a HV cable
with interlock (see parts that can be ordered).
Correct connection of masses, of the HV cable between the controller and pump
and of the interlock cable is illustrated in the "Ground Connections" figure.
When using the interlock cable furnished in the accessories bag, make sure that
CAUTION
no parts drop accidentally inside the controller.
158
Agilent IPCMini Ion Pump Controller User Manual